IONICRF™ GENERATOR

product photo

 

THE IONICRF™ GENERATOR

Careful thought is on display throughout every aspect of the IonicRF™ Generator's design and functionality. Its ergonomic form, intutitive user interface and streamlined navigation enables teams to quickly become proficient4, while the durable system architecture allows for reliable5 and high-efficiency performance6.

 

THERAPEUTIC VERSATILITY

The IonicRF™ Generator continues our evolution of RF therapy technologies with the capability to perform a range of procedure types, including:

  • Monopolar RF
  • Bipolar RF
  • Pulsed RF
  • Pulsed Dose RF
  • Simplicity™ Probe for SI Denervation

These comprehensive capabilities provide the flexibility needed to address chronic pain management across the spectrum of patient populations.

 

OPTIMAL PROCEDURE ROOM FOOTPRINT

Optimal procedure room footprint

The IonicRF™ Generator is lightweight and portable for a comfortable fit within standard procedure rooms. This system can be mounted on a pole with roll stand* for optimal maneuverability or set on a flat surface.

 

EFFICIENCY & RELIABILITY

The IonicRF™ Generator's architecture was designed to combine optimal system reliability5 with high-effiency preformance6.

The ProCharge™ intelligent power algorithm ensures efficient6 temperature ramping and ablation.

Independently controlled channels give the system the capacity to treat 4 sites at once with simultaneous and staggered start options1.

Real-time temperature monitoring results in consistent and predictable lesion formation1.


COMPLETE ACCESSORY COMPATIBILITY

The IonicRF™ Generator is compatible with Abbott’s existing line of needles and electrodes1.

Needles:1


16 G, 18 G, 20 G, 22 G

5 cm, 10 cm, 15 cm, 20 cm

Needle assortment is available in both the Classic and Select Lines. Premium Select Line needles are spray-coated and silicone-treated to: 

  • Substantially reduce penetration and running forces2
  • Maximize ease of use and patient comfort

Electrodes:1


  • Reusable and single-use
  • Offered in stainless steel or durable kink-resistant Nitinol

Simplicity™ Probe:1


Achieves effective sacroiliac denervation by creating a true strip lesion through a single insertion site providing greater anatomical coverage in less time3.


Read more about managing chronic pain and Abbott’s leading-edge portfolio.


RESOURCES AND DOCUMENTATION

Get access our product manuals and instructions for use.  Get Started >

Stay up to date with our latest product advisories.  Read More >

 

*Compatible pole mount bracket and pole available through Abbott. Only available in select geographies

TECH SPECS

Reorder Number Description

RFG-IONIC

IonicRF™ Generator

 

ACCESSORIES

Reorder Number Description

Bundle RF-POLE

Pole Mount Bracket and Pole

GP202D-AC

Grounding Pad

 

QUESTIONS ABOUT REIMBURSEMENT?

We’ve got the answers you’re looking for.

 

References
  1. Abbott IonicRF™ Generator Clinician’s Manual, Plano, TX, 2020.
  2. NeuroTherm Inc. RF SL/SMK Silicone Concentration Validation. Doc. VR-083. 2012.
  3. Schmidt, P.C., Pino, C. A., & Vorenkamp, K.E. (2014). Sacroiliac Joint Radiofrequency Ablation with a Multilesion Probe: A Case Series of 60 Patients (v1.0). Anesthesia-Analgesia.com, 119(2): 460-462.
  4. Abbott IonicRF™ Generator Summative Validation Report (90624300). 2020.
  5. Abbott IonicRF™ Generator Service Life Report test data on file (90263325). 2020.  
  6. Abbott IonicRF™ Generator System Compatibility DVTR  (90619520). 2020.         
IMPORTANT SAFETY INFORMATION

IonicRF™ Generator

Prescription and Safety Information

Read this section to gather important prescription and safety information.

Intended Use

The IonicRF™ Generator, in combination with approved compatible electrodes and cannulae, is intended for lesioning of neural tissue in the nervous system as an aid in the management of pain. 

Indications for Use

The IonicRF™ Generator, in combination with approved compatible electrodes and cannulae, is indicated as an aid in the management of pain in the nervous system. Examples include facet denervation, trigeminal rhizotomy, and related functional neurosurgical procedures.

Contraindications 

The use of this device is contraindicated in patients with systemic infection or local infection in the area of the procedure.

Warnings 

The following warnings apply to this generator.

Instructions for use. Read and understand the instructions for use provided in this clinician's manual before operating the generator. 

Hazardous electrical output. The generator is for use only by qualified medical personnel. 

Electric shock hazard. This device presents an electric shock hazard under certain conditions. Physicians need to be aware of the following warnings: 

  • To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. 
  • Only approved medical grade power cords can be used with the generator. Use only the power cord specified for this unit. No modification of this equipment is allowed. 
  • Do not attempt to service or modify the equipment. For service, contact Abbott Medical. 
  • Do not under any circumstances perform testing or maintenance on the equipment while it is being used on a patient. 
  • Replace the power cord or plug immediately if it is cracked, frayed, broken, or otherwise damaged. 
  • Turn off the equipment and unplug the power cord before cleaning or servicing. 
  • Do not allow any fluid to enter the ventilation holes or sockets. 

Equipment failure. A failure of the equipment could result in an unintended increase of output power. If unexpected parameters are observed that do not correspond to the preset values, halt the procedure immediately by pressing the Emergency Stop button on the top of the generator. Do not operate the equipment again until the source of the problem is identified and corrected. 

Explosion hazard. Do not use this equipment in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide. 

Fire hazard. This device presents a fire hazard under certain conditions. Physicians need to be aware of the following warnings: 

  • Do not use extension cords or adapters of any type. The power cord and plug must be intact and undamaged. 
  • Use recommended non-flammable agents for cleaning and disinfection whenever possible. See “Cleaning the Generator” (page 32). 
  • Flammable agents for cleaning, disinfecting, or as solvents of adhesives must be allowed to evaporate before using radiofrequency (RF) surgery.
  • Do not place electrodes near or in contact with flammable materials.

Pooling hazard. Flammable solutions may pool under the patient or in body depressions, such as the umbilicus, and in body cavities, such as the vagina. Remove fluids pooled in body depressions and cavities before using the generator. 

Ignition hazard. Be aware of and take care to avoid the danger of ignition of endogenous gasses (e.g., cotton and gauze saturated with oxygen may be ignited by sparks produced during normal use of the generator).

Risk of RF burns and unintended stimulation. Do not turn the generator power on while touching any electrodes or probes. Do not place a generator-connected electrode that is not being used in contact with the patient. 

Risk of RF burns to patient. This device presents a risk of RF burns to the patient under certain conditions. Physicians need to be aware of the following warnings: 

  • Ensure the patient does not come into contact with metal parts of the table and its accessories. Antistatic sheeting is recommended. 
  • Avoid skin-to-skin contact between different parts of patient’s body (for example between the arms and the body of the patient). Use dry gauze if necessary. 
  • Avoid using physiological monitoring equipment during a procedure. If monitoring is required, place monitoring electrodes as far as possible from the electrode. Monitoring devices that use needle electrodes are not recommended. In all cases, monitoring systems with high-frequency current limiting devices are recommended.
  • Connection of a patient to high-frequency surgical equipment and to electromyograph or evoked response equipment simultaneously may result in burns at the site of the electrodes and possible damage to the applied parts.
  • Position all cables to the electrode and grounding pad (also known as the return electrode, dispersive electrode, or neutral electrode) to avoid contact with other electrodes and other metal objects. 
  • Place temporarily unused electrodes connected to the generator in a container or area that is electrically isolated from the patient. Never place a generator-connected electrode that is not being used in contact with the patient. When not in use, place accessories in a clean, dry, highly visible area away from patient contact.
  • When not in use, place accessories in a clean, dry, highly visible area away from patient contact.
  • Only use grounding pads listed in the Accessories section of this manual. See "Appendix B: System Components and Accessories" (page 1). 
  • When positioning the grounding pad, select a well-vascularized muscular site with proximity to the procedure. See "Applying the Grounding Pad" (page 20). 
  • Place the long side of the grounding pad perpendicular to the direction of the current flow from the operative site. See "Applying the Grounding Pad" (page 20).
  • Do not place the grounding pad over scars, bony prominences, prostheses, hair, or EKG electrodes.
  • Do not place the grounding pad in a location where fluids may pool.
  • If the patient is sedated, place your hand on the backside of the grounding pad, while still leaving it attached to the patient. If the grounding pad is unreasonably hot (a temperature greater than 46°C), stop the procedure by pressing the Emergency Stop button.
  • If the patient complains of heating at the grounding pad site, stop the procedure and remove the grounding pad from the patient.
  • Before applying power to the electrodes, ensure that the entire area of the grounding pad is firmly adhered to a suitably prepared and appropriate area of the patient’s body as defined in the grounding pad instructions for use.
  • If the continuity monitor alarm is triggered, remove and discard the grounding pad. Place a new grounding pad on a fresh patch of skin.
  • If unexpected parameter readings are observed that do not correspond to the preset values, halt the procedure immediately by pressing the Emergency Stop button. Do not operate the equipment again until the source of the problem is identified and corrected.

Interference with active implants. Check whether the patient has a cardiac pacemaker or other active implantable device and, if so, obtain qualified advice before using the generator. Operating the generator may interfere with or damage the implanted device. 

Redirection of high-frequency currents. Check whether the patient has an electrically conductive implant and, if so, obtain qualified advice before using the generator. Operating the generator may cause concentration or redirection of high-frequency currents. 

Interference with other equipment. During RF lesioning procedures, the radiated electrical fields may interfere with other electrical medical equipment. See "Minimizing Electromagnetic Interference" (page 5). 

Shortwave or microwave equipment. Operation in close proximity to shortwave or microwave therapy equipment may produce instability in the applied parts. 

Apparent low output or failure of equipment. If low output is observed or the equipment does not function correctly at normal operating setting, check the grounding pad and its connections. 

Risk of patient injury. Do not use endoscopically. The accessories are not appropriate for endoscopic use. 

Proper device use. Do not operate the generator if the alert tones are not audible after the volume is adjusted. 

Non-sterile. The generator is non-sterile and should be kept outside of the sterile field. 

Accessories. Use only accessories approved by Abbott Medical. See “Appendix B: System Components and Accessories” (page 40). 

Continuity monitoring. The generator uses continuity monitoring. Loss of full contact between the neutral electrode and the patient will result in an auditory alarm.

Precautions 

The following precautions apply to this generator. 

Inspection. Inspect the generator and reusable accessories before each use. In particular, check the electrode cable insulation for possible damage. 

Mechanical damage. If the equipment has suffered any mechanical damage, return it to the supplier for inspection and testing before further use. 

Electrode positioning. Do not activate the generator output until the electrodes are correctly positioned in the patient. 

Use of fluids. If fluids are being used during a procedure, ensure that they are positioned away from the generator. 

Dispersive connections. Ensure that the grounding pad is connected to the patient and to the generator. 

Cleaning the generator. When cleaning the outer casing or touchscreen, do not use abrasive agents or solvents. See "Cleaning the Generator" (page 32). 

Emergency stop. For safety, always have someone positioned next to the Emergency Stop button during operation. If at any time the device is behaving erratically, press the Emergency Stop button, located on top of the generator, which will return the device to a safe state. For example, if the displayed temperature and graph do not match the desired set temperature.

Adverse Events 

Possible adverse events that may result from the use of this device include, but are not limited to, the following: 

  • As a consequence of electrosurgery, damage to surrounding tissue through iatrogenic injury can occur. 
  • Nerve injury, including thermal injury, or puncture of the spinal cord or nerve roots, potentially resulting in radiculopathy, paresis, and paralysis. 
  • Pain, pulmonary embolism, hemothorax or pneumothorax, infection, unintended puncture wound, including vascular puncture and dural tear, hemorrhage, and hematoma.


MAT-2006978 v1.0

 

MAT-2007075 v1.0

You are about to exit the Abbott family of websites for a 3rd party website

Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott.


The website that you have requested also may not be optimized for your screen size.

Do you wish to continue and exit this website?

test

True
accessibility

© 2018 Abbott. All Rights Reserved. Please read the Legal Notice for further details.

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company. 

accessibility

test

accessibility

You are about to exit the Abbott family of websites for a 3rd party website

Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott.


The website that you have requested also may not be optimized for your screen size.

Do you wish to continue and exit this website?